作者: Jacqueline Souza , Eunice Kazue Kano , Eunice Emiko Mori Koono , Simone Grigoleto Schramm , Valentina Porta
DOI: 10.1365/S10337-008-0913-Y
关键词:
摘要: This study describes an accurate, sensitive, and specific chromatographic method for the simultaneous quantitative determination of lamivudine zidovudine in human blood plasma, using stavudine as internal standard. The separation was performed a C8 column (150 × 4.6 mm, 5 μm), ultraviolet absorbency detection at 270 nm with gradient elution. Two mobile phases were used. Phase A contained 10 mM potassium phosphate 3% acetonitrile, whereas B methanol. linear used variability A-B phase proportion from 98–2% to 72–28%, respectively. drug extraction two 4 mL aliquots ethyl acetate.