作者: Min-Seong Choi , Jae-Chern Yoo
DOI: 10.3807/JOSK.2013.17.4.296
关键词:
摘要: A low density lipoprotein (LDL) cholesterol testing device, structured with serial reflected face-to-face mirror (SRM) allowing spectrophotometry measurements, is presented. The has been employed to measure the amount of light that a sample absorbs, but it generally should have had path length longer than 10 mm secure enough sensitivity. Such requirement often problematic in implementing thin type lab on disc (LOD). We developed SRM system which was implemented detection chamber 1.4mm thickness, providing 10mm, and thus straightforwardly being applicable LOD as compact disc. experimental results show gives not only much thinner design compared conventional spectrophotometry-based also comparable performance already commercialized spectrometers.