作者: Liyun Bai , Ping Ji , Xian Li , Hui Gao , Linlin Li
DOI: 10.1155/2019/4231872
关键词:
摘要: Individualized titanium mesh holds many advantages over conventional mesh. There are few reports in the literature about effect of pore size and thickness on mechanical properties This study is designed to develop an individualized using computer-assisted design additive manufacturing technology. will also explore different thicknesses sizes its through 3D FEA. According this study, increased when decreased (0.5 mm 0.3 mm). With increase diameter (3 mm 5 mm), decreased. The has less influence than does Titanium with a 0.4 mm strong enough causes stimulation mucosa; therefore, it more suitable for clinical use. In addition, parameters should be decided clinically according bone defect size, location, force situation.