[Can 99mTc-mercaptoacetyltriglycine (99mTc-MAG3) evaluate the renal function without blood sampling?: consensus report from multicenter study].

作者: E Tsukamoto , K Itoh , A Kubo , T Yamazaki , K Takahashi

DOI:

关键词:

摘要: A multicenter study was undertaken in Japan to evaluate the correlation between percentage of renal uptake 99mTc-mercaptoacetyltriglycine (99mTc-MAG3) estimated by count-based gamma camera method and blood clearance 99mTc-MAG3. Twenty four centers were enrolled 172 cases finally analyzed this study. The 99mTc-MAG3 (TER) obtained using a single sample taken at 44 min after injection. Comparison TER provided coefficient 0.874; suggesting that sufficiently accurate quantification function could be from estimate method. This also showed comparison might feasible among patients under same protocols, although precise careful consideration is required each center.

参考文章(0)