作者: Michael J. Flynn , Robert McGee , Joseph Blechinger
DOI: 10.1117/12.713805
关键词: Image resolution 、 Transverse plane 、 Imaging phantom 、 Flat panel detector 、 Spatial frequency 、 Tomosynthesis 、 Optics 、 Physics 、 Resolution (electron density) 、 Sagittal plane
摘要: We consider in this report, a tomosynthesis (TS) system utilizing recently developed Selenium flat panel detector (Shimadzu Safire II) capable of acquiring 30 pulsed frames per second (fps) with precise geometric registration to moving gantry. The reconstructed TS resolution achieved is reported relation the obtained modern volumetric computed tomography (vCT) systems. Computed assessed using sphere phantom. To describe coronal/sagittal re-formatted planes, we average transverse and axial response express typical vCT performance as 1.00 mm (FWHM PlSF) 0.83 cycles/mm (10% MTF). Tomosynthesis wire images, use FWHM LSF frequency at 20% peak spatial 2.41 LSF) 2.62 (20% Peak). Cadaver images confirm that about 3 times better than for On other hand, slice sensitivity larger LSF).