The use of direct sample insertion into a nitrogen-argon inductively coupled plasma for emission spectrometry-I: Technique optimization and application to the analysis of aluminium oxide

作者: G. Zaray , J.A.C. Broekaert , F. Leis

DOI: 10.1016/0584-8547(88)80050-8

关键词: ArgonAluminium oxidesAnalytical chemistryDetection limitAluminium oxideGraphiteChemistryCopperCalibration curveInductively coupled plasma

摘要: Abstract The determination of trace elements in aluminium oxide with the aid direct sample insertion inductively coupled plasma atomic emission spectrometry (ICP-AES) has been studied. form and shape graphite cups containing as well gas flows other ICP parameters were optimized dry solution residues zinc, copper strontium. It was found that after mixing proportion 1:1 powder briquetting pellets 10–20 mg, relatively volatile such Cd, Cu, Mg Zn could be volatilized detected down to μg g level. For titanium, which forms refractory carbides, effect different volatilization aids (NH4C1, AgCl, BaF2 PTFE) With addition PTFE quantitative determinations are possible a detection limit 7 is obtained. Relative standard deviations case peak areas obtained from time-resolved simultaneous measurements line background intensities at order 5 % calibration curves linear over least one magnitude.

参考文章(25)
Eric D. Salin, R. L. A. Sing, Microsample liquid analysis with a wire-loop direct sample insertion technique in an inductively coupled plasma Analytical Chemistry. ,vol. 56, pp. 2596- 2598 ,(1984) , 10.1021/AC00277A074
Jose A. C. Broekaert, Brigette. Wopenka, Hans. Puxbaum, Inductively coupled plasma optical emission spectrometry for the analysis of aerosol samples collected by cascade impactors Analytical Chemistry. ,vol. 54, pp. 2174- 2179 ,(1982) , 10.1021/AC00250A010
Youbin Shao, Gary Horlick, Performance of a Direct Sample Insertion System for the Inductively Coupled Plasma Applied Spectroscopy. ,vol. 40, pp. 386- 393 ,(1986) , 10.1366/0003702864509178
Brian R. Eggins, Norman H. Smith, Diffusion control in linear sweep voltammetry Analytical Chemistry. ,vol. 51, pp. 2282- 2283 ,(1979) , 10.1021/AC50049A050
Zhang Li-Xing, G. F. Kirkbright, M. J. Cope, J. M. Watson, A Microprocessor-Controlled Graphite Rod Direct Sample Insertion Device for Inductively Coupled Plasma Optical Emission Spectrometry Applied Spectroscopy. ,vol. 37, pp. 250- 254 ,(1983) , 10.1366/0003702834634569
Michael Thompson, John E. Goulter, Friedrich Sieper, Laser ablation for the introduction of solid samples into an inductively coupled plasma for atomic-emission spectrometry Analyst. ,vol. 106, pp. 32- 39 ,(1981) , 10.1039/AN9810600032