An integrated head-holder/coil for intraoperative MRI in open neurosurgery.

作者: Andreas Staubert , Otto Pastyr , Gernod Echner , Arnulf Oppelt , Theodor Vetter

DOI: 10.1002/(SICI)1522-2586(200005)11:5<564::AID-JMRI13>3.0.CO;2-N

关键词: Head holderSurface coilElectromagnetic coilNeurosurgeryRadiologyMagnetic resonance imagingMedicineTumor surgeryIntraoperative MRIImage quality

摘要: With the invention of "open" magnetic resonance imaging (MRI) systems, access to patient is possible during procedure. An important application these systems intraoperative MRI control extent resection tumor surgery. Up now flexible surface coils wrapped around, or placed at each side head, were used for imaging. These have several disadvantages such as unreliability, interindividual problems, difficult handling, poor hygienic properties, and often unsatisfactory inhomogeneous image quality. To solve most an MR-compatible head-holder in combination with integrated coil use a 0.2 T C-shaped magnet was developed. Forty-eight patients known cranial tumors underwent intraoperatively. In 32 (67%), residual found, additional surgical performed. The head-holder/coil safe practical tool MRI, providing efficient reliable neurosurgical procedures.

参考文章(4)
Rainer Wirtz, Volker Tronnier, Friedrich Albert, Michael Knauth, Mario Bonsanto, Andreas Staubert, Otto Pastyr, Stefan Kunze, Modified Headholder and operating table for intra-operative MRI in neurosurgery. Neurological Research. ,vol. 20, pp. 658- 661 ,(1998) , 10.1080/01616412.1998.11740580
Volker M. Tronnier, Christian R. Wirtz, Michael Knauth, Gerald Lenz, Otto Pastyr, Mario M. Bonsanto, Friedrich K. Albert, Rainer Kuth, Andreas Staubert, Wolfgang Schlegel, Klaus Sartor, Stefan Kunze, Intraoperative diagnostic and interventional magnetic resonance imaging in neurosurgery. Neurosurgery. ,vol. 40, pp. 891- 902 ,(1997) , 10.1097/00006123-199705000-00001
Garnette R. Sutherland, Taro Kaibara, Deon Louw, John Saunders, A mobile high-field magnetic resonance imaging system for neurosurgery Neurosurgical Focus. ,vol. 6, ,(1999) , 10.3171/FOC.1999.6.3.9
Peter McL. Black, Thomas Moriarty, Eben Alexander, Philip Stieg, Eric J. Woodard, P. Langham Gleason, Claudia H. Martin, Ron Kikinis, Richard B. Schwartz, Ferenc A. Jolesz, Development and implementation of intraoperative magnetic resonance imaging and its neurosurgical applications. Neurosurgery. ,vol. 41, pp. 831- 845 ,(1997) , 10.1097/00006123-199710000-00013