作者: Sophie Gillette-Guyonnet , Sandrine Andrieu , Fati Nourhashemi , Christelle Cantet , Hélène Grandjean
DOI: 10.1016/S0047-6374(02)00199-9
关键词: Bone mineral 、 Dual-energy X-ray absorptiometry 、 Bone mineral content 、 Measurement method 、 Soft tissue mass 、 Total fat 、 Nuclear medicine 、 Medicine 、 Whole body 、 Anthropometry 、 Developmental biology 、 Ageing
摘要: Information is lacking regarding the comparison of bone mineral and body composition measurements obtained from different dual energy X-ray absorptiometry (DXA) systems. The purpose this study to compare whole density (BMD) hologic QDR-4500 W (DXA H) lunar L) Seven women older than 40 years age, were scanned on same day by using DXA H L observer. Measured total fat mass, percentage fat, free soft tissue mass BMD similar between machines, but was significantly higher with (P=0.0156). This due a content L. Agreement two systems quantified Bland Altman method. For each variable, all pairs within 95% confidence limits. These results suggested that we could use measurement methods interchangeably. However, inter-machine differences, comparisons machines should be interpreted caution.