Evaluation of differences between fan-beam and pencil-beam densitometers.

作者: R. B. Mazess , H. S. Barden

DOI: 10.1007/S002230001147

关键词:

摘要: We compared the bone and body composition results in vivo on two densitometers using fan-beam geometry (EXPERT PRODIGY) with those pencil-beam (DPX). Measurements were made large groups of adults ranging weight from about 50 to 120 kg. Both spine femur neck BMD averaged within 1% results, there was no magnitude dependence by Bland-Altman analysis. Total BMC PRODIGY DPX congruent, but EXPERT, 2% lower higher than corresponding values DPX. Soft-tissue closely congruent for DPX; comparable EXPERT-DPX differences showed greater scatter significant dependence. The smaller fan-angle (4°) probably contributed its better congruence EXPERT (12°).

参考文章(14)
Richard Mazess, Charles H. Chesnut, Michael McClung, Harry Genant, Enhanced precision with dual-energy x-ray absorptiometry Calcified Tissue International. ,vol. 51, pp. 14- 17 ,(1992) , 10.1007/BF00296209
G M Blake, J C Parker, F M A Buxton, I Fogelman, Dual X-ray absorptiometry: a comparison between fan beam and pencil beam scans British Journal of Radiology. ,vol. 66, pp. 902- 906 ,(1993) , 10.1259/0007-1285-66-790-902
H. Franck, M. Munz, M. Scherrer, Bone Mineral Density of Opposing Hips Using Dual Energy X-Ray Absorptiometry in Single-Beam and Fan-Beam Design Calcified Tissue International. ,vol. 61, pp. 445- 447 ,(1997) , 10.1007/S002239900365
T. Fuerst, H. K. Genant, Evaluation of body composition and total body bone mass with the hologic QDR-4500 Osteoporosis International. ,vol. 6, pp. 202- 202 ,(1996) , 10.1007/BF02500344
S E Bouyoucef, I D Cullum, P J Ell, Cross-calibration of a fan-beam X-ray densitometer with a pencil-beam system. British Journal of Radiology. ,vol. 69, pp. 522- 531 ,(1996) , 10.1259/0007-1285-69-822-522
J. Martin Bland, DouglasG. Altman, Statistical methods for assessing agreement between two methods of clinical measurement. The Lancet. ,vol. 327, pp. 307- 310 ,(1986) , 10.1016/S0140-6736(86)90837-8
N. A. Pocock, K. A. Noakes, Y. Majerovic, M. R. Griffiths, Magnification error of femoral geometry using fan beam densitometers Calcified Tissue International. ,vol. 60, pp. 8- 10 ,(1997) , 10.1007/S002239900177
N Barthe, P Braillon, D Ducassou, B Basse-Cathalinat, Comparison of two Hologic DXA systems (QDR 1000 and QDR 4500/A). British Journal of Radiology. ,vol. 70, pp. 728- 739 ,(1997) , 10.1259/BJR.70.835.9245885
R. B. Mazess, J. A. Hanson, R. Payne, R. Nord, M. Wilson, Axial and total-body bone densitometry using a narrow-angle fan-beam. Osteoporosis International. ,vol. 11, pp. 158- 166 ,(2000) , 10.1007/PL00004178