Dual X-ray absorptiometry: a comparison between fan beam and pencil beam scans

作者: G M Blake , J C Parker , F M A Buxton , I Fogelman

DOI: 10.1259/0007-1285-66-790-902

关键词:

摘要: The recent introduction of dual X-ray absorptiometry (DXA) systems with fan beam instead conventional pencil scanning geometry represents a significant technical advance in bone densitometry. This report describes phantom and in-vivo studies the effect change configuration on DXA measurements. Fan measurements acquired one new generation scanners, Hologic QDR-2000, were compared scans performed an earlier model, QDR-1000. variation height above table anthropomorphic spine were: mineral content (BMC): -3.1% cm-1; projected area (AREA): -2.8% density (BMD): -0.2% cm-1. For magnitude was less than 0.1% QDR-2000 scan results for 20 volunteers correlated closely QDR-1000 data (r = 0.966-0.998). BMD hip, differences between consistent errors expected from measurement precision. AREA BMC data, however, larger observed individual deviations which body habitus subjects. Although to significantly affected clinically more important negligibly small.

参考文章(9)
Donald F. Morrison, Multivariate statistical methods Published in <b>1976</b> in New York NY) by McGraw-Hill. ,(1976)
P Tothill, Methods of bone mineral measurement. Physics in Medicine and Biology. ,vol. 34, pp. 543- 572 ,(1989) , 10.1088/0031-9155/34/5/001
I. D. Cullum, P. J. Ell, J. P. Ryder, X-ray dual-photon absorptiometry: a new method for the measurement of bone density. British Journal of Radiology. ,vol. 62, pp. 587- 592 ,(1989) , 10.1259/0007-1285-62-739-587
HEINZ W. WAHNER, WILLIAM L. DUNN, MANUEL L. BROWN, RICHARD L. MORIN, B. LAWRENCE RIGGS, Comparison of dual-energy x-ray absorptiometry and dual photon absorptiometry for bone mineral measurements of the lumbar spine. Mayo Clin., Proc.; (United States). ,vol. 63, pp. 1075- 1084 ,(1988) , 10.1016/S0025-6196(12)65502-5
Richard Mazess, Bruce Collick, Joel Trempe, Howard Barden, James Hanson, Performance evaluation of a dual-energy X-ray bone densitometer Calcified Tissue International. ,vol. 44, pp. 228- 232 ,(1989) , 10.1007/BF02556569
T. L. KELLY, D. M. SLOVIK, D. A. SCHOENFELD, R. M. NEER, Quantitative digital radiography versus dual photon absorptiometry of the lumbar spine. The Journal of Clinical Endocrinology and Metabolism. ,vol. 67, pp. 839- 844 ,(1988) , 10.1210/JCEM-67-4-839
F M A Buxton, G M Blake, J C Parker, M K Lewis, I Fogelman, Short communication: spinal bone density measured with a 5 s scan. British Journal of Radiology. ,vol. 66, pp. 275- 277 ,(1993) , 10.1259/0007-1285-66-783-275
James A. Sorenson, Peter R. Duke, Steven W. Smith, Simulation studies of dual-energy x-ray absorptiometry Medical Physics. ,vol. 16, pp. 75- 80 ,(1989) , 10.1118/1.596405
G. M. Blake, C. M. Tong, I. Fogelman, Intersite comparison of the Hologic QDR-1000 dual energy X-ray bone densitometer. British Journal of Radiology. ,vol. 64, pp. 440- 446 ,(1991) , 10.1259/0007-1285-64-761-440