作者: K W Marich , W J Treytl , J G Hawley , N A Peppers , R E Myers
DOI: 10.1088/0022-3735/7/10/015
关键词:
摘要: Instrumental improvements in the laser microprobe system designed earlier have been developed. The chief are (i) redesigned components cavity for greater stability, (ii) binocular viewing microscope equipped with optional phase optics and camera, incorporating safety devices to protect viewer condenser from beam, (iii) direct reading energy monitor pulse indicator show whether one or more giant pulses were delivered, (iv) interchangeable lens mirror light-coupling systems special uses, (v) retroalignment precise optical alignment of sample plume spectrograph, (vi) six-tube polychromator multielement analysis.