Clinical use of a new position-independent rebound tonometer.

作者: Kerstin Stephanie Jablonski , André Rosentreter , Stergiani Gaki , Alexandra Lappas , Thomas Stefan Dietlein

DOI: 10.1097/IJG.0B013E318259AA47

关键词:

摘要: Purpose:To compare intraocular pressure (IOP) measurements obtained by rebound tonometry (Icare PRO tonometer), applanation (Goldmann and Perkins tonometry), dynamic contour in the upright supine positions, to investigate influence of axial length central corneal thickness.Methods:Ninety-nine right eyes 99 patients with glaucoma or suspect for glaucoma, admitted our department between November 2010 January 2011 obtain an IOP profile including measurements, were included study. position using Icare (RTPRO), a Goldmann (GAT), Pascal tonometer (DCT). In position, taken RTPRO hand-held (PAT). The means SDs all tonometers compared. Agreement was calculated Bland-Altman method.Results:The mean IOPs 17.78.0 mm Hg 17.6 +/- 7.8 19.9 6.6 Correlation analysis these data indicated good correlation readings GAT (r=0.951; P<0.001), DCT (r=0.897; P<0.001). revealed differences (bias) GAT, 0.1 -1.8 Hg, 95% limits agreement -3.6 3.8 -7.3 3.6 respectively. 19.2 6.4 19.6 6.2 PAT.Conclusions:Measurements RTPRO, either show those provided tonometry. study registered DRKS (German Clinical Trials Register; http://www.germanctr.de; DRKS00000581).

参考文章(11)
H A Quigley, Number of people with glaucoma worldwide. British Journal of Ophthalmology. ,vol. 80, pp. 389- 393 ,(1996) , 10.1136/BJO.80.5.389
Antti Kontiola, A new electromechanical method for measuring intraocular pressure. Documenta Ophthalmologica. ,vol. 93, pp. 265- 276 ,(1997) , 10.1007/BF02569066
G�nter K. Krieglstein, Wolfgang K. Waller, Goldmann applanation versus hand-applanation and .schiötz indentation tonometry. Graefes Archive for Clinical and Experimental Ophthalmology. ,vol. 194, pp. 11- 16 ,(1975) , 10.1007/BF00408271
J. Martin Bland, DouglasG. Altman, Statistical methods for assessing agreement between two methods of clinical measurement. The Lancet. ,vol. 327, pp. 307- 310 ,(1986) , 10.1016/S0140-6736(86)90837-8
E. S. Perkins, Hand-held applanation tonometer. British Journal of Ophthalmology. ,vol. 49, pp. 591- 593 ,(1965) , 10.1136/BJO.49.11.591
Charles D. Phelps, Gary K. Phelps, Measurement of intraocular pressure: a study of its reproducibility. Graefes Archive for Clinical and Experimental Ophthalmology. ,vol. 198, pp. 39- 43 ,(1976) , 10.1007/BF00411443
NIELS EHLERS, THORKILD BRAMSEN, STEFFEN SPERLING, APPLANATION TONOMETRY AND CENTRAL CORNEAL THICKNESS Acta Ophthalmologica. ,vol. 53, pp. 34- 43 ,(2009) , 10.1111/J.1755-3768.1975.TB01135.X
Liane H. van der Jagt, Nomdo M. Jansonius, Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry*. Ophthalmic and Physiological Optics. ,vol. 25, pp. 429- 435 ,(2005) , 10.1111/J.1475-1313.2005.00318.X