In vitro evaluation of the acetabular cup primary stability by impact analysis.

作者: Adrien Michel , Romain Bosc , Romain Vayron , Guillaume Haiat

DOI: 10.1115/1.4029505

关键词:

摘要: The implant primary stability of the acetabular cup (AC) is an important parameter for surgical success press-fit procedures used insertion cementless hip prostheses. In previous studies by our group (Mathieu, V., Michel, A., Lachaniette, C. H. F., Poignard, Hernigou, P., Allain, J., and Haiat, G., 2013, “Variation Impact Duration During in vitro Insertion Acetabular Cup Implants,” Med. Eng. Phys., 35(11), pp. 1558–1563) (Michel, Bosc, R., Mathieu, 2014, “Monitoring Press-Fit Measurements: Influence Bone Abrasion,” Proc. Inst. Mech. Eng., Part H, 228(10), 1027–1034), impact momentum duration were shown to carry information on AC within bone tissue. aim present study relate recorded during biomechanical stability. experimental protocol consisted testing 13 bovine samples that underwent successively series 15 reproducible mass falls impacts (5 kg, 5 cm) followed tangential testing. Each sample was tested with different hole sizes order obtain configurations. reach a maximum value interference fit equal around 1 mm. Moreover, correlation between obtained all configuration (R2 = 0.65). can be assessed through measurement force signal analysis. This opens new paths development medical device which could as decision support system assist surgeon implant.

参考文章(43)
Rina Sakai, Aki Kikuchi, Towa Morita, Naonobu Takahira, Katsufumi Uchiyama, Takeaki Yamamoto, Mistutoshi Moriya, Kentaro Uchida, Kensuke Fukushima, Kensei Tanaka, Masashi Takaso, Moritoshi Itoman, Kiyoshi Mabuchi, Hammering sound frequency analysis and prevention of intraoperative periprosthetic fractures during total hip arthroplasty. Hip International. ,vol. 21, pp. 718- 723 ,(2011) , 10.5301/HIP.2011.8823
Thomas P. Sculco, Mathias Bostrom, Timothy Wright, Khaled J. Saleh, Brian Bear, Initial stability of press-fit acetabular components: an in vitro biomechanical study. American journal of orthopedics. ,vol. 37, pp. 519- 522 ,(2008)
Phillip Cornwell, S Giardini, R Meneghini, None, Monitoring femoral component installation using vibration testing. Biomedical sciences instrumentation. ,vol. 41, pp. 13- 18 ,(2005)
Vincent Mathieu, Adrien Michel, Charles-Henri Flouzat Lachaniette, Alexandre Poignard, Philippe Hernigou, Jérome Allain, Guillaume Haïat, Variation of the impact duration during the in vitro insertion of acetabular cup implants. Medical Engineering & Physics. ,vol. 35, pp. 1558- 1563 ,(2013) , 10.1016/J.MEDENGPHY.2013.04.005
ERWIN MORSCHER, HEINZ BEREITER, CHRISTOPH LAMPERT, Cementless press-fit cup. Principles, experimental data, and three-year follow-up study. Clinical Orthopaedics and Related Research. ,vol. 249, pp. 12- 20 ,(1989) , 10.1097/00003086-198912000-00003
Paul G. Perona, Jeffrey Lawrence, Wayne G. Paprosky, Avinash G. Patwardhan, Mark Sartori, Acetabular micromotion as a measure of initial implant stability in primary hip arthroplasty Journal of Arthroplasty. ,vol. 7, pp. 537- 547 ,(1992) , 10.1016/S0883-5403(06)80076-8
Manal M. Shalabi, Johannes G. C. Wolke, Vincent M. J. I. Cuijpers, John A. Jansen, Evaluation of bone response to titanium-coated polymethyl methacrylate resin (PMMA) implants by X-ray tomography Journal of Materials Science: Materials in Medicine. ,vol. 18, pp. 2033- 2039 ,(2007) , 10.1007/S10856-007-3160-0
R. M. PILLIAR, J. M. LEE, C. MANIATOPOULOS, Observations on the effect of movement on bone ingrowth into porous-surfaced implants Clinical Orthopaedics and Related Research. ,vol. 208, pp. 108- 113 ,(1986) , 10.1097/00003086-198607000-00023
E. Adler, S.A. Stuchin, F.J. Kummer, Stability of press-fit acetabular cups. Journal of Arthroplasty. ,vol. 7, pp. 295- 301 ,(1992) , 10.1016/0883-5403(92)90052-R
STEPHEN D. COOK, KIMBERLY A. WALSH, RAY J. HADDAD, Interface mechanics and bone growth into porous Co-Cr-Mo alloy implants. Clinical Orthopaedics and Related Research. pp. 271- ,(1985) , 10.1097/00003086-198503000-00037