摘要: A shielded gradient system and radio frequency probes for a 4.7-T/30-cm horizontal bore magnet are described. The systems were designed principally to obtain images localized spectra from rat brains. gradients can operate in either or non mode. rf probe, which combines both highest sensitivity resolution, consists of large modified saddle type transmitter small surface reception coil, actively decoupled using diodes. performance is demonstrated on phantoms brain, where an image with plane resolution 80μ m slice thickness 150 μm obtained. © 1992 Academic Press, Inc.

参考文章(7)
Y Zur, S Stokar, A phase-cycling technique for canceling spurious echoes in NMR imaging Journal of Magnetic Resonance. ,vol. 71, pp. 212- 228 ,(1987) , 10.1016/0022-2364(87)90051-5
R Turner, B Chapman, A.M Howseman, R.J Ordidge, R Coxon, P Glover, P Mansfield, Snap-shot magnetic resonance imaging at 0.1 T using double-screened gradients Journal of Magnetic Resonance. ,vol. 80, pp. 248- 258 ,(1988) , 10.1016/0022-2364(88)90294-6
M.Robin Bendall, David T Pegg, Uniform sample excitation with surface coils for in vivo spectroscopy by adiabatic rapid half passage Journal of Magnetic Resonance. ,vol. 67, pp. 376- 381 ,(1986) , 10.1016/0022-2364(86)90447-6
R S Parker, I Zupancic, J Pirs, Coil system to produce orthogonal, linear magnetic field gradients Journal of Physics E: Scientific Instruments. ,vol. 6, pp. 899- 900 ,(1973) , 10.1088/0022-3735/6/9/028
P Mansfield, B Chapman, Active magnetic screening of coils for static and time-dependent magnetic field generation in NMR imaging Journal of Physics E: Scientific Instruments. ,vol. 19, pp. 540- 545 ,(1986) , 10.1088/0022-3735/19/7/008
Cecil E Hayes, William A Edelstein, John F Schenck, Otward M Mueller, Matthew Eash, An efficient, highly homogeneous radiofrequency coil for whole-body NMR imaging at 1.5 T Journal of Magnetic Resonance (1969). ,vol. 63, pp. 622- 628 ,(1985) , 10.1016/0022-2364(85)90257-4
P.A Bottomley, H.R Hart, W.A Edelstein, J.F Schenck, L.S Smith, W.M Leue, O.M Mueller, R.W Redington, NMR imaging/spectroscopy system to study both anatomy and metabolism. The Lancet. ,vol. 322, pp. 273- 274 ,(1983) , 10.1016/S0140-6736(83)90250-7