In vitro behaviour of mechanical heart valves in a 1.5 T superconducting magnet

作者: H. Frank , P. Buxbaum , L. Huber , S. Globits , D. Glogar

DOI: 10.1007/BF00187551

关键词:

摘要: During in vitro tests, we exposed seven mechanical cardiac valves to a 1.5 Tesla magnetic field and measured their deflection temperature changes during gradient-echo sequence. We used special suspension system which indicated the largest at entrance tube any deflections within gradient field. A sensor valve ring both air water bath exposure high-frequency The Bjork-Shiley (Shiley Irvine, CA, USA) displayed most coil, namely 1.8 mm. All other models less deflection. No took place sequences. Three an increase temperature. Medtronic with 0.6°C.

参考文章(5)
R. Maas, H. Kooijman, M. Heller, H. Langkowski, E. Grabbe, E. Bücheler, Risks and artifacts in MRT (1.5 tesla) caused by metallic foreign bodies. Studies on phantoms and 60 patients Rofo-fortschritte Auf Dem Gebiet Der Rontgenstrahlen Und Der Bildgebenden Verfahren. ,vol. 147, pp. 365- 374 ,(1987) , 10.1055/S-2008-1048659