Cross-calibration of DXA equipment: Upgrading from a hologic QDR 1000/W to a QDR 2000

作者: Kenneth G. Faulkner , Claus-C. Gl�er , Michelle Estilo , Harry K. Genant

DOI: 10.1007/BF00308312

关键词:

摘要: In this study, the cross-calibration of a fan beam DXA system (Hologic QDR-2000) to pencil scanner from same manufacturer QDR-1000/W) is described. The scanners were calibrated by using anthropomorphic spine phantom at installation. To verify consistent machine calibration, group 69 female subjects, aged 46–75, had anteroposterior (AP) and proximal femur scans on QDR-1000/W followed array sites QDR-2000 during visit. Many subjects bilateral examinations for total 123 hip scans. Pencil area, bone mineral content (BMC), density (BMD) compared with values obtained QDR-1000/W, linear regression equations derived relating two instruments. At spine, no differences found between BMD values. A slight difference modes was detected but not deemed clinically significant. Linear models AP measurements showed correlation coefficients greater than 0.99, slopes 1.00, intercepts equivalent zero, small root mean square errors. Comparisons equivalency femoral neck trochanter regions machines in mode, increases BMC other both mode. Correlation 0.97–0.99 all measurement except Ward's. Regression 1.00–1.04, significantly different zero residual As enough that they clinical significance. However, longitudinal drug trials requiring highly precise determination spinal changes, these may be important.

参考文章(5)
C C Glüer, P Steiger, R Selvidge, K Elliesen-Kliefoth, C Hayashi, H K Genant, Comparative assessment of dual-photon absorptiometry and dual-energy radiography. Radiology. ,vol. 174, pp. 223- 228 ,(1990) , 10.1148/RADIOLOGY.174.1.2294552
J. Lilley, B. G. Walters, D. A. Heath, Z. Drolc, In vivo and in vitro precision for bone density measured by dual-energy X-ray absorption. Osteoporosis International. ,vol. 1, pp. 141- 146 ,(1991) , 10.1007/BF01625443
DJ Sartoris, D Resnick, Dual-energy radiographic absorptiometry for bone densitometry: current status and perspective. American Journal of Roentgenology. ,vol. 152, pp. 241- 246 ,(1989) , 10.2214/AJR.152.2.241
Von Stetten, Cross-calibration of dual-energy X-ray absorptiometry (DXA) bone densitometry systems Calcified Tissue International. ,vol. 52, pp. 166- ,(1993)