作者: V DOROSCHUK
DOI: 10.1016/J.TALANTA.2004.03.056
关键词:
摘要: Abstract The possibility to use 4-(2-pyridylazo)resorcinol (PAR) and 1-(2-pyridylazo)-2-naphthol (PAN) for manganese(II) concentrating by the micellar extraction at cloud point (CP) temperature subsequent atomic absorption spectrometry (AAS) determination was investigated. Under optimum conditions, preconcentration of 100 ml water sample in presence 1% non-ionic surfactant (NS) OP-7, 1×10 −4 M permitted detection 5 μg l −1 manganese. proposed method has been applied AAS manganese samples after extraction.