Real-time proton beam range monitoring by means of prompt-gamma detection with a collimated camera

作者: F Roellinghoff , A Benilov , D Dauvergne , G Dedes , N Freud

DOI: 10.1088/0031-9155/59/5/1327

关键词: Bragg peakAtomic numberBeam (structure)Collimated lightProtonNeutronOpticsDetectorPhysicsRange (particle radiation)Radiological and Ultrasound TechnologyRadiology Nuclear Medicine and imaging

摘要: Prompt-gamma profile was measured at WPE-Essen using 160 MeV protons impinging a movable PMMA target. A single collimated detector used with time-of-flight (TOF) to reduce the background due neutrons. The target entrance rise and Bragg peak falloff retrieval precision determined as function of incident proton number by fitting procedure independent data sets. Assuming improved sensitivity this camera design greater detectors, precisions 1 2 mm (rms) are expected for clinical pencil beam. TOF improves contrast-to-noise ratio performance method significantly.

参考文章(22)
Jong-Won Kim, Pinhole Camera Measurements of Prompt Gamma-rays for Detection of Beam Range Variation in Proton Therapy Journal of the Korean Physical Society. ,vol. 55, pp. 1673- 1676 ,(2009) , 10.3938/JKPS.55.1673
Han Rim Lee, Jong Hoon Park, Chan Hyeong Kim, Chul Hee Min, Design optimization of a 2D prompt-gamma measurement system for proton dose verification Journal of the Korean Physical Society. ,vol. 61, pp. 239- 242 ,(2012) , 10.3938/JKPS.61.239
M Moteabbed, S España, H Paganetti, Monte Carlo patient study on the comparison of prompt gamma and PET imaging for range verification in proton therapy Physics in Medicine and Biology. ,vol. 56, pp. 1063- 1082 ,(2011) , 10.1088/0031-9155/56/4/012
E. Sterpin, J. Sorriaux, S. Vynckier, Extension of PENELOPE to protons: Simulation of nuclear reactions and benchmark with Geant4 Medical Physics. ,vol. 40, pp. 111705- ,(2013) , 10.1118/1.4823469
Harald Paganetti, Range uncertainties in proton therapy and the role of Monte Carlo simulations Physics in Medicine and Biology. ,vol. 57, ,(2012) , 10.1088/0031-9155/57/11/R99
Kim gi Dong, Mono-Energetic Neutron Source by 7 Li(p,n) 7 Be Reaction Journal of the Korean Physical Society. ,vol. 55, pp. 1404- 1408 ,(2009) , 10.3938/JKPS.55.1404
Fine Fiedler, Georgy Shakirin, Judith Skowron, Henning Braess, Paulo Crespo, Daniela Kunath, Jörg Pawelke, Falk Pönisch, Wolfgang Enghardt, On the effectiveness of ion range determination from in-beam PET data. Physics in Medicine and Biology. ,vol. 55, pp. 1989- 1998 ,(2010) , 10.1088/0031-9155/55/7/013
J. E. Gillam, I. Torres-Espallardo, C. Lacasta, G. Llosa, J. Barrio, V. Stankova, C. Solaz, M. Rafecas, Hodoscope coincidence imaging for hadron therapy using a compton camera ieee nuclear science symposium. pp. 3508- 3513 ,(2011) , 10.1109/NSSMIC.2011.6152645
L Grevillot, D Bertrand, F Dessy, N Freud, D Sarrut, GATE as a GEANT4-based Monte Carlo platform for the evaluation of proton pencil beam scanning treatment plans Physics in Medicine and Biology. ,vol. 57, pp. 4223- 4244 ,(2012) , 10.1088/0031-9155/57/13/4223
M. Testa, M. Bajard, M. Chevallier, D. Dauvergne, N. Freud, P. Henriquet, S. Karkar, F. Le Foulher, J. M. Létang, R. Plescak, C. Ray, M.-H. Richard, D. Schardt, E. Testa, Real-time monitoring of the Bragg-peak position in ion therapy by means of single photon detection. Radiation and Environmental Biophysics. ,vol. 49, pp. 337- 343 ,(2010) , 10.1007/S00411-010-0276-2