作者: F.E. Maly , A. Urwyler , H.P. Rolli , C.A. Dahinden , A.L. de Weck
DOI: 10.1016/0003-2697(88)90344-2
关键词: Optics 、 Linearity 、 Photon 、 Camera lens 、 Luminescence 、 Analytical chemistry 、 Image resolution 、 Detector 、 Microtiter plate 、 Chemistry 、 Luminescent Measurements 、 Biophysics 、 Cell biology 、 Biochemistry 、 Molecular biology
摘要: A combination of a two-dimensional photon detector (double-microchannel plate) with single-photon sensitivity and an optical projection system that allows space-resolved quantitation luminescent emissions from spatially extended objects is described. "luminescent image" the object focused onto accumulated over preset time stored in digital frame memory which counts areas interest can be read. In this study, consisted microtiter plate containing samples was placed below projecting lens (2.0/21 mm, 36 X 24-mm format camera lens) at distance 38.5 cm. Although geometry substantially limited collection, achieved only 10X less than obtained dedicated photon-counting luminometer. slightly diminished collection peripheral wells apparently caused by could corrected arithmetically. Both chemically generated luminescence (ATP bioluminescence) cell-derived, superoxide-dependent (with lucigenin as chemilumigenic probe) were detected excellent spatial resolution linearity response wide range.