Determination of heparin levels in blood with activated partial thromboplastin time by a piezoelectric quartz crystal sensor

作者: Tzong-Jih Cheng , Tsun-Mei Lin , Tsui-Hsun Wu , Hsien-Chang Chang

DOI: 10.1016/S0003-2670(00)01346-5

关键词:

摘要: Abstract A piezoelectric quartz crystal (PQC) sensor, possessing a sensitive frequency response to viscoelastic variations of substance loading on an electrode surface, was used determine both whole blood activated partial thromboplastin time (WBaPTT) and plasma (PLaPTT) induced by anticoagulant heparin. The PQC sensor results showed linear relationship between WBaPTT (or PLaPTT) heparin levels in clinically relevant concentrations (0–0.4 unit/ml). Mean individual R2 (= 0.9491) for regressive curve adequately shown. method can be evaluate the effect through determination owing comparable sensitivity (P

参考文章(19)
G.J. Despotis, S.A. Santoro, E. Spitznagel, K.M. Kater, J.L. Cox, P. Barnes, D.G. Lappas, Prospective evaluation and clinical utility of on-site monitoring of coagulation in patients undergoing cardiac operation. The Journal of Thoracic and Cardiovascular Surgery. ,vol. 107, pp. 271- 279 ,(1994) , 10.1016/S0022-5223(94)70481-3
H. Muramatsu, E. Tamiya, M. Suzuki, I. Karube, Quartz-crystal gelation detector for the determination of fibrinogen concentration Analytica Chimica Acta. ,vol. 217, pp. 321- 326 ,(1989) , 10.1016/S0003-2670(00)80413-4
H. Muramatsu, K. Kimura, T. Ataka, R. Homma, Y. Miura, I. Karube, A quartz crystal viscosity sensor for monitoring coagulation reaction and its application to a multichannel coagulation detector. Biosensors and Bioelectronics. ,vol. 6, pp. 353- 358 ,(1991) , 10.1016/0956-5663(91)85022-O
Hans K Pilegaard, J.Michael Hasenkam, Lars Knudsen, Hans Henrik Kimose, Hanne Grønnesby, Jørn Attermann, Niels Trolle Andersen, Practicability of patient self-testing of oral anticoagulant therapy by the international normalized ratio (INR) using a portable whole blood monitor. A pilot investigation. Thrombosis Research. ,vol. 85, pp. 77- 82 ,(1997) , 10.1016/S0049-3848(96)00224-1
Hsien-Chang Chang, Tzong-Jih Cheng, Tsui-Hsun Wu, Tsun-Mei Lin, Determination of coagulation time in whole blood containing anticoagulant by piezoelectric quartz crystal sensor Sensors and Actuators B-chemical. ,vol. 66, pp. 296- 298 ,(2000) , 10.1016/S0925-4005(99)00462-1
G. L. Hayward, J. M. Scharer, R. L. Dutton, Z. Zhang, M. Moo Young, Thickness shear mode acoustic assay for plasminogen activators Analytical Communications. ,vol. 35, pp. 25- 28 ,(1998) , 10.1039/A707538C
George J. Despotis, Jr Hogue, Sam A. Santoro, J. Heinrich Joist, Patrick W. MT Barnes, Demetrios G. Lappas, Effect of heparin on whole blood activated partial thromboplastin time using a portable, whole blood coagulation monitor. Critical Care Medicine. ,vol. 23, pp. 1674- 1679 ,(1995) , 10.1097/00003246-199510000-00011
K Keiji Kanazawa, Joseph G Gordon, The oscillation frequency of a quartz resonator in contact with liquid Analytica Chimica Acta. ,vol. 175, pp. 99- 105 ,(1985) , 10.1016/S0003-2670(00)82721-X
Si Shi-Hui, Zhou Tie-an, Liu De-Zhong, Nie Li-Hua, Yao Shou-Zhuo, Using Piezoelectric Quartz Crystal to Study Hemorheological Phenomena: Blood Clotting and Urokinase Activated Fibrinolysis Analytical Letters. ,vol. 27, pp. 2027- 2037 ,(1994) , 10.1080/00032719408007232
Si Shi Hui, Shen Da Zhong, Nie Li Hua, Yao Shou Zhuo, Impedance study of erythrocyte aggregation and sedimentation by piezoelectric crystal sensor Bioelectrochemistry and Bioenergetics. ,vol. 36, pp. 161- 164 ,(1995) , 10.1016/0302-4598(94)01762-P