Icare rebound tonometry in children with known and suspected glaucoma

作者: Meghan S. Flemmons , Ya-Chuan Hsiao , Jacqueline Dzau , Sanjay Asrani , Sarah Jones

DOI: 10.1016/J.JAAPOS.2010.11.022

关键词:

摘要: Background Accurate intraocular pressure (IOP) measurement, important in managing pediatric glaucoma, often presents challenges. The Icare rebound tonometer shows promise for screening healthy children and has been reported comparable with Goldmann applanation adults glaucoma. purpose of this study was to evaluate the against clinic IOP measurement Methods This a prospective comparing versus tonometry Children known or suspected glaucoma were recruited from scheduled visits. measured by clinician subsequently (GAT) different single masked clinician. Results A total 71 eyes included. GAT ranged 9 36 mm Hg. readings 11 44 Mean difference between 2.3 ± SD 3.7 Hg, p Conclusions within 3 Hg 63% greater than 75%. device may be reasonable estimate selected whose cannot otherwise obtained clinic; however, correlation IOPs clinical findings must continue considered each case.

参考文章(17)
W. Schreiber, C.K. Vorwerk, A. Langenbucher, W. Behrens-Baumann, A. Viestenz, Evaluierung der Reboundtonometrie (ICare) im Vergleich zum TonopenXL und dem Goldmann-Applanationstonometer Ophthalmologe. ,vol. 104, pp. 299- 304 ,(2007) , 10.1007/S00347-007-1487-Z
Leon N. Davies, Hannah Bartlett, Edward A. H. Mallen, James S. Wolffsohn, Clinical evaluation of rebound tonometer Acta Ophthalmologica Scandinavica. ,vol. 84, pp. 206- 209 ,(2006) , 10.1111/J.1600-0420.2005.00610.X
Nima Pakrou, Tim Gray, Richard Mills, John Landers, Jamie Craig, Clinical Comparison of the Icare Tonometer and Goldmann Applanation Tonometry Journal of Glaucoma. ,vol. 17, pp. 43- 47 ,(2008) , 10.1097/IJG.0B013E318133FB32
Makoto Nakamura, Urtogtah Darhad, Yasuko Tatsumi, Miyuki Fujioka, Azusa Kusuhara, Hidetaka Maeda, Akira Negi, Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers. American Journal of Ophthalmology. ,vol. 142, pp. 332- 334 ,(2006) , 10.1016/J.AJO.2006.02.035
Paolo Brusini, Maria Letizia Salvetat, Marco Zeppieri, Claudia Tosoni, Lucia Parisi, Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients. Journal of Glaucoma. ,vol. 15, pp. 213- 217 ,(2006) , 10.1097/01.IJG.0000212208.87523.66
CARLOS GARC??A-RES??A, JOS?? M. GONZ??LEZ-MEIJOME, JORGE GILINO, EVA YEBRA-PIMENTEL, Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes. Optometry and Vision Science. ,vol. 83, pp. 102- 107 ,(2006) , 10.1097/01.OPX.0000200673.96758.7B
Jose M. Martinez-de-la-Casa, Julian Garcia-Feijoo, Federico Saenz-Frances, Gianmarco Vizzeri, Ana Fernandez-Vidal, Carmen Mendez-Hernandez, Julian Garcia-Sanchez, Comparison of rebound tonometer and Goldmann handheld applanation tonometer in congenital glaucoma. Journal of Glaucoma. ,vol. 18, pp. 49- 52 ,(2009) , 10.1097/IJG.0B013E31816F760C
S. Munkwitz, A. Elkarmouty, E. M. Hoffmann, N. Pfeiffer, H. Thieme, Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range Graefes Archive for Clinical and Experimental Ophthalmology. ,vol. 246, pp. 875- 879 ,(2008) , 10.1007/S00417-007-0758-3
P. Fernandes, J. A. Diaz-Rey, A. Queiros, J. M. Gonzalez-Meijome, J. Jorge, Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population Ophthalmic and Physiological Optics. ,vol. 25, pp. 436- 440 ,(2005) , 10.1111/J.1475-1313.2005.00327.X