Evaluation of column hardware on liquid chromatography-mass spectrometry of phosphorylated compounds.

作者: Hiroshi Sakamaki , Takeharu Uchida , Lee Wah Lim , Toyohide Takeuchi

DOI: 10.1016/J.CHROMA.2014.12.088

关键词:

摘要: Abstract The influences of column hardware, such as chromatographic tubes and frits, on liquid chromatography–mass spectrometry (LC–MS) analysis phosphorylated compounds were evaluated. signal to noise ratio (S/N) the intensity flavin adenine dinucleotide (FAD) using a glass lined tube polyethylene frit (GL–PE) was approximately 170 90 times higher, respectively, than those conventional stainless steel (S–S) column. In addition, retention time FAD GL–PE shortest compared other columns. Interaction between metal ions in flow path S–S stronger that them Thus, low. Since specific surface area pair frits 70 larger (150 mm × 2.1 mm), found have more effective improvement S/N well tubes, when analyzed by LC–MS. When evaluated LC–MS(/MS) column, increased.

参考文章(29)
Quoc-Chon Le, Marie-Hélène Ropers, Hélène Terrisse, Bernard Humbert, Interactions between phospholipids and titanium dioxide particles. Colloids and Surfaces B: Biointerfaces. ,vol. 123, pp. 150- 157 ,(2014) , 10.1016/J.COLSURFB.2014.09.010
Akira Wakamatsu, Kentaro Morimoto, Masao Shimizu, Shinobu Kudoh, A severe peak tailing of phosphate compounds caused by interaction with stainless steel used for liquid chromatography and electrospray mass spectrometry Journal of Separation Science. ,vol. 28, pp. 1823- 1830 ,(2005) , 10.1002/JSSC.200400027
Dominic Winter, Joerg Seidler, Yael Ziv, Yosef Shiloh, Wolf D. Lehmann, Citrate boosts the performance of phosphopeptide analysis by UPLC-ESI-MS/MS. Journal of Proteome Research. ,vol. 8, pp. 418- 424 ,(2009) , 10.1021/PR800304N
Yoshiki Asakawa, Noriaki Tokida, Chinami Ozawa, Masahiko Ishiba, Osamu Tagaya, Naoki Asakawa, Suppression effects of carbonate on the interaction between stainless steel and phosphate groups of phosphate compounds in high-performance liquid chromatography and electrospray ionization mass spectrometry Journal of Chromatography A. ,vol. 1198-1199, pp. 80- 86 ,(2008) , 10.1016/J.CHROMA.2008.05.015
Toyohide Takeuchi, Daido Ishii, High-performance micro packed flexible columns in liquid chromatography Journal of Chromatography A. ,vol. 213, pp. 25- 32 ,(1981) , 10.1016/S0021-9673(00)80628-7
Thomas De Vijlder, Jasper Boschmans, Erwin Witters, Filip Lemière, Study on the loss of nucleoside mono-, di- and triphosphates and phosphorylated peptides to a metal-free LC–MS hardware International Journal of Mass Spectrometry. ,vol. 304, pp. 83- 90 ,(2011) , 10.1016/J.IJMS.2010.05.022
Klaus K. Unger, Romas Skudas, Michael M. Schulte, Particle packed columns and monolithic columns in high-performance liquid chromatography-comparison and critical appraisal. Journal of Chromatography A. ,vol. 1184, pp. 393- 415 ,(2008) , 10.1016/J.CHROMA.2007.11.118
Stephen Brunauer, P. H. Emmett, Edward Teller, Adsorption of Gases in Multimolecular Layers Journal of the American Chemical Society. ,vol. 60, pp. 309- 319 ,(1938) , 10.1021/JA01269A023